findmanualpdf.com

Gastec 13TP Carbon Disulphide Detector Tube Instructions

Quick guide for Gastec 13TP Carbon Disulphide Detector Tube. Includes measurement procedures, sampling rates, interference data, and safety precautions for accurate gas detection.

Table of contents

Quick guide from the manual

The Gastec 13TP Detector Tube is designed for measuring Carbon Disulphide in air or industrial atmospheric conditions. This guide outlines the necessary steps for operation using a compatible Gastec Air Sampling Pump (Model GSP-300FT-2 or GSP-501FT). Always ensure the tube is used within the specified temperature (0-35°C) and humidity (0-90% RH) ranges for accurate results.

Measurement Procedure

  1. Confirm the black colour inlet rubber tube holder is equipped with the pump.
  2. Break the tips off both ends of a fresh primary and secondary tube using the pump's tube tip breaker.
  3. Connect the ends marked with the circle symbol using rubber tubing.
  4. Insert the tube into the pump inlet, ensuring the arrow on the tube points toward the pump.
  5. Set the flow rate to 100 mL/min and the timer to 10 minutes on the pump.
  6. Press the start switch to begin sampling.
  7. Once sampling is complete, remove the tube and read the concentration level where the stained reagent meets the unstained reagent.
  8. If the reading is higher than 1.0 ppm, repeat the process using a fresh tube with a flow rate of 50 mL/min and a timer set to 10 minutes.
  9. Apply correction factors for sampling rate and atmospheric pressure if necessary.

Interferences

The following substances may interfere with the reading. If these gases are present, the test result may be affected:

  • Hydrogen sulphide: ≧ 0.1 ppm (Interference: +)
  • Carbonyl sulphide: ≧ 0.8 ppm (Interference: +)
  • Dimethyl disulphide: ≧ 1.0 ppm (Interference: +)
  • Sulphur dioxide: ≧ 0.1 ppm (Interference: +)

Methylene Chloride, Acetone, and Methanol do not cause discolouration and do not interfere within their specified ranges.

Environmental Conditions and Corrections

No temperature or humidity correction is required within the range of 0-35°C and 0-90% R.H. For pressure corrections, use the following formula: Tube Reading (ppm) × 1013 (hPa) / Atmospheric Pressure (hPa).

Safety and Disposal

Caution: When breaking tube ends, keep away from eyes. Do not touch broken glass or reagents with bare hands. The pretreatment tube contains a small amount of hexavalent chromium. Dispose of all tubes according to local government rules and regulations.

Manufacturer information

Gastec Corporation

Brand profile

Practical help

Common problems

Reading higher than 1.0 ppm

Prepare a fresh tube, set the flow meter to 50 mL/min, and set the timer to 10 minutes to re-sample.

Interference from coexisting gases

Be aware that Hydrogen sulphide, Carbonyl sulphide, Dimethyl disulphide, and Sulphur dioxide can cause positive interference.

Need for pressure correction

Use the formula: Tube Reading (ppm) × 1013 (hPa) / Atmospheric Pressure (hPa).

Before use

  • Confirm the black colour inlet rubber tube holder is equipped with the pump.
  • Ensure the pump is a compatible Gastec model (GSP-300FT-2 or GSP-501FT).
  • Check the expiration date on the box of tubes.
  • Verify the ambient temperature is between 0-35°C.
  • Verify the relative humidity is between 0-90%.

Specs in practice

Measuring Range
The concentration range the tube can detect (0.05 - 1.0 ppm or 1.0 - 2.4 ppm depending on settings).
Sampling Rate
The required airflow speed through the tube (100 mL/min or 50 mL/min).
Detecting Limit
The lowest concentration detectable by the tube (0.04 ppm).

Model compatibility

  • Compatible with Gastec Gas Sampling device Model GSP-300FT-2 or GSP-501FT.

Manual page author

Michael Turner

Technical manual editor

Reviews PDF manuals for structure, safety notes, and practical product details so readers can find the right information quickly.