findmanualpdf.com

Nonin 1000USB-C USB Interface Adapter

The Nonin 1000USB-C interface adapter enables real-time data transfer from Nonin handheld pulse oximeters to Continua-compliant Application Hosting Devices (AHD) or PCs.

Table of contents

Product Overview

The Nonin 1000USB-C interface adapter is designed to connect Nonin handheld pulse oximeters to a Continua-compliant Application Hosting Device (AHD) or a personal computer (PC). This adapter facilitates the transfer of real-time data between the pulse oximeter and the host device, provided the host supports the ISO/IEEE 11073-10404:2010 standard for pulse oximeters.

Safety Information

To prevent the risk of electric shock to the patient or operator, follow the setup sequence strictly. Do not touch the patient and the accessible pins on the accessory simultaneously. Ensure all connected equipment complies with relevant IEC standards (e.g., IEC 60950 for data processing, IEC 60601-1 for medical equipment). Only use Nonin-branded accessories to avoid improper performance or patient injury.

Setup and Installation

Follow these steps to install the adapter:

  • Follow the instructions provided by your Continua AHD vendor to set up the host device.
  • Connect the pulse oximeter sensor connector to the Oximeter adapter port.
  • Connect the sensor to the Sensor adapter port.
  • Connect the adapter port to the USB port on your Continua AHD or PC.
  • Follow the AHD vendor instructions for data collection.

The LED on the unit will blink green until the appropriate PHDC driver is installed and loaded. Once the drivers are loaded and the unit is ready to transfer data, the LED will light steady green.

System Integration

Incorporating the device into a medical system requires the integrator to identify, analyze, and evaluate risks to patients, operators, and third parties. Any subsequent changes to the medical system, such as changing the configuration or adding/disconnecting devices, may introduce new risks and require additional analysis. Issues resulting from user-initiated changes may include data corruption or loss.

Care and Maintenance

Clean the device by wiping it with a soft cloth dampened with a mild detergent or a 10% bleach solution. Do not use undiluted bleach or other cleaning solutions, as this may cause permanent damage. Dry the device with a soft cloth or allow it to air dry.

Practical help

Common problems

LED does not turn steady green

Ensure the appropriate PHDC driver is installed and loaded on your Continua AHD or PC.

Data transfer issues

Verify that the host device is Continua Certified and supports the required ISO/IEEE standard.

Before use

  • Verify compatibility of the monitor, sensor, and accessories.
  • Ensure all equipment is suitable for the patient's environment.
  • Check that the host device is properly set up according to the vendor's instructions.
  • Ensure the device is clean and free of damage.

Specs in practice

Operating Temperature
-20 °C to 50 °C (-4 °F to 122 °F)
Operating Humidity
10% to 90% non-condensing
Storage Temperature
-40 °C to 70 °C (-40 °F to 158 °F)

Model compatibility

  • Designed for use with all Nonin handheld pulse oximeters.
  • Supports real-time data collection only; memory playback is not supported.
  • The AHD vendor is responsible for the AHD, application, and USB PHDC host driver.

Manual page author

David Miller

Documentation analyst

Organizes user manual content into clear summaries, with attention to model details, product context, and everyday usability.