Steri-Run Decontamination Instructions for Thermo Fisher Scientific Vios 160i and i160 Incubator

Quick guide for running the Steri-Run decontamination routine on the Thermo Fisher Scientific Vios 160i and i160 incubators, including preparation, operation, and cancellation steps.

Table of contents

Quick guide for Steri-Run decontamination

This document provides instructions for running the Steri-Run decontamination routine on the Heracell Vios 160i and i160 incubators.

Preparation

Before starting the routine, ensure the device is prepared:

  • Remove all samples and accessories (external probes, Hepa Filter, etc.) from the workspace. Note: The Unit Temp sensor must remain in the unit.
  • Drain water from the reservoir and wipe up any residual water.
  • Remove the silicone plug from the inner chamber side and plug it into the outer side of the access port to prevent damage.

Starting the Steri-Run

  1. Press the Steri-Run key to access the menu.
  2. Press the START key to activate the routine. The Steri-Run Progress dialog window will appear.
  3. Open the glass door until the buzzer sounds (approximately 30 seconds).
  4. Close the door after the buzzer sounds.

Operation and Completion

The decontamination routine will begin, and the device will heat up. The electromechanical door lock will engage at 65°C. The display will show the current status, including temperature, start time, phase, and remaining run time.

The routine takes approximately 12 hours to complete. Once finished, open both device doors to air the workspace when the audible alarm sounds.

Canceling the routine

The Steri-Run decontamination routine can be interrupted at any time by pressing the Stop key. If the Stop key is pressed, the Steri-Run stop dialog box will be displayed for a safety scan.

Practical help

Common problems

Routine needs to be stopped

Press the Stop key to interrupt the routine and initiate a safety scan.

Door lock engagement

The electromechanical door lock engages automatically at 65°C during the routine.

Before use

  • Remove all samples and accessories
  • Keep the Unit Temp sensor in the unit
  • Drain water from the reservoir
  • Wipe up residual water
  • Move silicone plug to the outer side of the access port

Specs in practice

Steri-Run duration
Approximately 12 hours
Door lock temperature
Engages at 65°C

Model compatibility

  • Applies to Heracell Vios 160i and i160 models.